Basic Info.
Model NO.
Drawell-Emerald 800T/H
Certification
CE, ISO9001
Analysis Method
End-Point, Fixed-Time, Rate
Wavelength Range
16 Wavelengths From 340 to 800nm
Sample Volume
2UL ~70UL (0.1UL Stepping)
Transport Package
Standard Export Package with Pure English
Specification
1508x865.5x1204mm
Product Description
Drawell-Emerald Auto Biochemistry Analyzer (800T/H) Parameter:
1. 800 test per hour,160 cuvettes
2. 180 sample position.
3. 200 refrigerated reagent position.
4. optical grating
Features:
1. Fully automated, Discrete, Random access,STAT priority,real-time dual wavelength.
2. Total open setting of items, supported different reagents from different countries.
3. Reversed optics and imported photo sensor make sure the precision of detection results.
4. Microsring injector imported from Germany assures the veracity of a trace of samples,
sample step by 0.1 UL, reagent step by0.5 UL
5. 24 hour non-stop working, convenient emergency function with inserting emergency at random.
6. Implementing analysis of ultra-trace reaction solution to 100 UL, and saving more reagents.
7. Discrete cuvette, preventing from cross-infection, and be washed and reused, making sure accurate result and save cost.
8. Perfect self-detection function makes sure the analyzer in order and accurate result.
9. Software is easy to learn and use. And test report can be printed.
10. Perfect function of quality control management is convenient for users to manage the clinical test result.
11. Detecting in the whole reaction process with reaction curve, convenient for users to monitor thoroughly and scientifically analyze the detection data.
12. Supporting software can implement the function of detecting surplus reagent through liquid level induction of reagent probe.
Overall Performance | Model | Drawell-EMERALD |
Type | Discrete, Fully automatic |
Throughput | 800T/H |
Analysis Method | End-point, Fixed-time,Rate,etc |
Programming Items | programmable for any test items |
Certification | CE, FDA, ISO9001, ISO14001, ISO13485 |
Sample unit | Sample Position | 180 samples, double disk design, including 60 samples for calibration/quality dribbling refrigerated |
Sample Tubes | Original blood collection tube,micro tube,plastic tube,etc. |
Sample Types | Serum,plasma,urine,etc. |
Sample Volume | 2ul ~70ul (0.1ul stepping) |
Sample Probe | Having liquid level induction function, ensuring the accuracy of adding sample |
Probe Cleaning | Automatic washing both interior and exterior , cross contamination rate < 0.1% |
Analysis Items | analyzing maximally 100 items at the same time |
Reagent unit | Reagent Position | 200 reagent positions, dual-tray design |
Wash Solution | Washing solution can be installed in any position, including needle, stir bar washing solution and water bath disinfectant |
Reagent Volume | 20-350ul, 0.5ul increment |
Temperature Control | continuous2~8ºCcooling within 24 hours |
Reagent Probe | Having liquid level induction function, ensuring the accuracy of adding reagent. Inventory checking |
Probe cleaning | Automatic washing both interior and exterior |
Reaction unit | Reaction Volume | 150-550ul | Reaction Temperature | 37ºC |
Temperature system | Water bath,Real-time temperature display |
Cuvettes | 160 positions with permanent hard glass cuvettes (optical path length 5mm) |
Stirrer | Independent stir bar, nanotechnology, highly polished, automatic frequency conversion |
Cleaning | automatic water cleaning and drying system |
System Adjustment | System can record the blank value and get more accurate result by deducting blank base value of each cuvette |
Optical System | Light Source | Halogen lamp, lamp can be auto sleep and wake up |
Receiver | Holographic flat field concave grating diode array detection |
Wavelength Range | 16 wavelengths from 340 to 800nm |
Linear Range | -0.3000~3.3000 Abs |
Stray light | < 0.1% or Abs > 3.0 |
Calibration and QC | Calibration Method | Linear calibration:K factor,two- and multiple linear Nonlinear Calibration:spline,polynomial,Logarithm,etc. |
Quality Control | Auto QC,cooled QC positions |
Software | Emergency Function | Emergency sample can be placed in the sample tray at any time |
Software | Windows XP (SP3) / Win 7 |
Online | Full bi-directional communication possible |
Data Processing | Having reaction curve detection and data display function in the whole process |
Others | Water consumption | Deionized water , <45 L/hour |
Power Supply | 110-240V ,50Hz ,<1.5KW |
Dimension | 1508×865.5×1204mm |
Weight | 400Kg |
Address:
Suite 1117, Lane561 Xiuchuan Rd., Greeland Max-Mall, Pudong New Area, Shanghai, China
Business Type:
Manufacturer/Factory, Trading Company, Group Corporation
Business Range:
Chemicals, Electrical & Electronics, Health & Medicine, Industrial Equipment & Components, Instruments & Meters, Manufacturing & Processing Machinery
Management System Certification:
ISO 9001, ISO 9000, ISO 14001, QSR
Company Introduction:
Drawell International Technology Co., Ltd. Is a professional scientific instrument manufacturer and supplier; We consistently provide the highest quality consumable lab equipment and best service for life technologies. In cooperation with many domestic enterprises we provide users with first-class products. The main products are UV/Visible Spectrophotometers, AAS, Gas Chromatography, Centrifuges, Electronic Balances, Microscopes, Electrochemical Instruments Incubators and other analytical instruments&&laboratory equipments. They have been widely used in different kinds of scopes, such as bioscience research, pharmacological science, quality control, food industry, environmentology and education. With a history of more than 20 years, DRAWELL is committed to provide high-quality scientific instruments for all the users around the world. In recent years, we have invested a lot to optimize our R&D.
With professional technical research and support team, we hope to make comprehensive business cooperation with analytic instrument manufacturers and distributors from all over the world.